94 Agilent 1200 Series Autosampler Reference Manual
3 Troubleshooting and Test Functions
Draw Metering device draws the defined
injection volume.
Command also positions the
vial at the seat, lifts the
needle, and lowers the needle
into vial. Command can be
done more than once
(maximum draw volume of
100µl cannot be exceeded).
Use “Plunger Home” to reset
the metering device.
Needle Up Lifts the needle out of the vial. Command also switches the
valve to bypass if it is not
already in that position.
Vial to Tray Returns the selected vial to the tray
position.
Command also lifts the needle
to the upper position.
Needle into Seat Lowers the needle arm into the seat. Command also returns the vial
to the tray position.
Mainpass Switches the injection valve to the
mainpass position.
Needle Up/Mainpass Lifts the needle arm to the upper
position and Switches the injection
valve to the mainpass position.
Command available from
Control Module only.
Table 11 Injector Step Commands (continued)
Step Action Comments

Table of Contents

Other manuals for Agilent Technologies G1314B VWD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1314B VWD and is the answer not in the manual?

Agilent Technologies G1314B VWD Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1314B VWD
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals