EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>G1364B

Agilent Technologies G1364B User Manual

Agilent Technologies G1364B
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
Agilent 1100 Series Fraction Collectors User’s Guide 25
Configuration and Operation of the Fraction Collector 1
Delay volumes and delay calibration
26 Save the method.
Figure 9 Fraction Collection Parameters dialog box with delay volume entry
NOTE
If you anticipate running samples at a variety of flow rates, it is best to calibrate at the
highest likely flow rate first and install the appropriate length tubing. This way, there will be
sufficient delay for the lower flow rates and new lengths will not need to be cut.
CAUTION
Before using the preparative scale fraction collector G1364B the delay calibration
adapter G1364-87301 has to be removed from the needle carrier assembly.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1364B and is the answer not in the manual?

Agilent Technologies G1364B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1364B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals