Agilent 1100 Series Fraction Collectors User’s Guide 97
Specifications 5
Performance Specifications for the Fraction Collectors
Delay volumes [µl] Fraction collector inlet to diverter valve: ~500 (typical, depends on length of
the tubing)
Diverter valve: ~15
Diverter valve to needle: ~110
Needle: ~5
Delay calibration
sensor
Single wavelength absorbance detector working at 654 nm, consisting of a
LED and a photo diode
Diverter valve 3/2 Diverter valve with low internal volume (15 µl), switching time
< 100 ms, maximum operating pressure 6 bar
cooling Optional (with additional G1330B), performance depending on ambient
conditions and the volume of collected fractions
maximum capacity 3 fraction collectors in parallel plus one recovery fraction collector
connected via 12-Position, 13-Port Selector valve (PN G1160A)
GLP features Early maintenance feedback (EMF), electronic records of maintenance and
errors
Interfaces Controller-area network (CAN).
optional; LAN or external contacts interface
RS232C,
APG-remote (for remote start / stop signals to / from other modules)
Interface to G1330A Thermostat
CAN-DC-out for operation of Agilent
approved external devices like valves
Safety features Leak detection and safe leak handling, error detection and display, exhaust
fan for fume extraction of hazardous vapors
* Vials can be used as recommended by Agilent Technologies (see “List of Recommended Vials and
Caps” on page 83 and “List of Recommended Plates and Closing Mats” on page 85) but must be
uncapped. Only the 96 deep well-plates can be used (without closing mats, see “List of Recom-
mended Plates and Closing Mats” on page 85)
Tabl e 23 performance Specifications Agilent 1100 Series PREPARATIVE SCALE Fraction
Collector (G1364B) (continued)
Type Specification
NOTE
Only one type of well-plates can be used at a time in one tray.