98 Agilent 1100 Series Fraction Collectors User’s Guide
5Specifications
Performance Specifications for the Fraction Collectors
Tabl e 24 Performance Specifications Agilent 1100 Series ANALYTICAL SCALE Fraction
Collector (G1364C)
Type Specification
trigger modes Time slices, Peak (threshold, up- / downslope), Timetable
(combination of time intervals and peak) and Manual trigger
(supported only with G1323B Control Module)
Agilent 1100 UV-Vis detectors (G1314A, G1315A/B, G1365 A/B and
the Agilent G1946C/D, G1956A/B LC-MSD are fully supported. Other
detectors can be used but are not supported for fraction collection
operating modes Discrete fractions: default mode for all vessels. The flow is diverted
to waste, while moving from one vessel position to the next vessel
position
Continuous flow: optional, available only when using the deep well
plates. It is possible to move from one well plate position to the next
one without diverting the flow into the well plate to waste
Needle into location: Needle pushes into the vessel as deep as
specified, for the use with capped vials and test tubes and well
plates with closing mats
Droplet setup mode: The tip of the fraction collector needle will
initially move down to the bottom of the well. Then it will slowly
move upwards while the fraction is collected. The droplet setup
mode enables the fraction collector to collect small fractions without
bubbles.
fraction vessel capacities
and trays
• 4 x well-plates full tray (MTP)
*
• 2 x well-plates std. tray + 10 funnels with external containers
*
(+ 1
half tray)
• 2 × well-plates std. tray (MTP) + 10 × 2 ml vials
*
(+ 1 half tray)
• 100 x 2 ml in std. tray (+ 1 half tray)
*
• 3 x 40 x 2 ml in half tray
*
• 3 x 40 funnels in half tray
• 3 x 15 x 6 ml in half tray
*
• Full tray with 40 test tubes (30 mm OD, max. height 48 mm, ~20 ml
vol.)
• Full tray with 60 test tubes (25 mm OD, max. height 48 mm)
• Full tray with 126 test tubes (16 mm OD, max. height 48 mm)
• Full tray with 215 test tubes (12 mm OD, max. height 48 mm)
Installed trays are automatically detected and identified. Installed
plates and vials can be detected when operating in the needle into
location mode
* max. height can be extended by using the short needle assembly
G1364-87202