7 Error Information
Pump Error Messages
1260 Infinity II Flexible Pump User Manual 132
Solvent counter exceeded limit
Error ID: 29146
The counter for the solvent volume has exceeded the limit, which has been set in
the user interface.
Waste counter limit exceeded
Error ID: 29147
The counter for the waste volume has exceeded the limit, which has been set in
the user interface.
• Parameter: None
Probable cause Suggested actions
1 No solvent present.
Refill solvent bottle.
2 Inappropriate setting.
Check solvent counter setting in user interface.
Probable cause Suggested actions
1 The waste container is full.
Empty waste container.
2 Inappropriate setting for waste counter.
• Reset waste counter.
• Adjust waste counter limit.