12 LAN Configuration
Manual Configuration
1260 Infinity II Flexible Pump User Manual 292
With Telnet
Whenever a TCP/IP connection to the module is possible (TCP/IP parameters set
by any method), the parameters may be altered by opening a Telnet session.
1 Open the system (DOS) prompt window by clicking on Windows START
button and select “Run...”. Type “cmd” and press OK.
2 Type the following at the system (DOS) prompt:
• c:\>telnet <IP address> or
• c:\>telnet <host name>
Figure 36 Telnet - Starting a session
where <IP address> may be the assigned address from a configuration
session with the Handheld Controller, or the default IP address (see
“Configuration Switch” on page 284).
When the connection was established successfully, the module responds
with the following:
Figure 37 A connection to the module is made