7 Error Information
Pump Error Messages
1260 Infinity II Flexible Pump User Manual 144
Timeout: Wait for Volume
Error ID: 29182
A target condition (volume, which is the delivered flow since the limit has been
set) has been sent to the instrument which should have been reached within an
expected time frame but didn’t. Either the limit, time frame or the current value of
the variable has been modified later directly or indirectly (for example the flow
rate).
Timeout: Wait for Flow
Error ID: 29183
A target condition (flow rate) has been sent to the instrument which should have
been reached within an expected time frame but didn’t. Either the limit, time
frame or the current value of the variable has been modified later directly or
indirectly.
Probable cause Suggested actions
1 Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Verify control software, macros, manual
commands.
Probable cause Suggested actions
1 Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Verify control software, macros, manual
commands.