1260 Infinity II Flexible Pump User Manual 135
7 Error Information
Pump Error Messages
Pump drive blocked or encoder failed
Error ID: 29214
Pump drive blocked or encoder failed.
• Parameter: None
Drive current too low
Error ID: 29205
The current consumption of the pump drive is too low.
• Parameter: 1 – 2 referring to pump drive
Probable cause Suggested actions
1 Blockage of the system before pressure
sensor.
Please contact your Agilent service
representative.
2 Drive encoder failed.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Drive motor defect.
Please contact your Agilent service
representative.
2 Wrong/missing connection of pump drive to
main board.
Please contact your Agilent service
representative.