7 Error Information
Pump Error Messages
1260 Infinity II Flexible Pump User Manual 140
Insufficient power of drive encoder LED
Error ID: 29235
Insufficient power of drive encoder LED.
• Parameter: 1 – 2 referring to pump drive
Drive encoder error
Error ID: 29237, 29238, 29239, 29215
An error has occurred for the pump drive encoder.
• Parameter: 1 – 2 referring to pump drive
Writing the multipurpose valve tag failed
Error ID: 29241
Writing the multipurpose valve tag failed.
• Parameter: None
Probable cause Suggested actions
1 Pump drive encoder is defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Pump drive encoder is defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Multipurpose valve head tag defect.
Replace multipurpose valve head.
2 Multipurpose valve tag head reader is defect.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7104C and is the answer not in the manual?

Agilent Technologies G7104C Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7104C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals