1260 Infinity II Flexible Pump User Manual 103
5 How to Optimize the Performance of Your Module
Using Solvent Calibration Tables
6 The imported solvent table will now appear in the Solvent Type Catalogs.
The imported solvent table is now available for selection in the Method
Settings of the pump module.
Recommendations for Setting up Methods Using
Salt-Containing Eluents
For LC Drivers 3.4 or newer, please use Ammonium Sulfate in Water V.04 and
Sodium Chloride in Water V.04 and manually input the molarity (0.5M-3.0M or
0.5M-2.0M respectively) as shown in the example below.
With the 1290 Infinity II Bio LC System Agilent offers a dedicated instrument for
running applications that are typical for the (Bio)pharma market, such as IEX (Ion
Exchange Chromatography) or HIC (Hydrophobic Interaction Chromatography),
and applications in Academia, Government and Applied Markets, using
concentrated salt solutions as eluents.
For LC Drivers 3.3 or older, please choose the solvent definition most suitable for
the application according to the chart below.
NOTE
Ammonium Sulphate V.03 and Sodium Chloride V.03 only properly characterize
the G7132A High-Speed Pump. Do not use these with any quaternary/flexible
pump configurations.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7104C and is the answer not in the manual?

Agilent Technologies G7104C Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7104C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals