1260 Infinity II Flexible Pump User Manual 141
7 Error Information
Pump Error Messages
Unknown multipurpose valve type
Error ID: 29242
The type information of the multipurpose valve is invalid.
• Parameter: None
Pump drive encoder error
Error ID: 29211
The pump drive encoder has generated no signal.
• Parameter: 1 – 2 referring to pump drive
Pump drive error
Error ID: 29212, 29213
The pump drive failed during calibration.
• Parameter: 1 – 2 referring to pump drive
Probable cause Suggested actions
1 Wrong valve head installed.
Check or replace multipurpose valve head.
2 Valve head has invalid RFID tag content.
Check or replace multipurpose valve head.
Probable cause Suggested actions
1 Pump drive encoder is defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Pump drive encoder is defect.
Please contact your Agilent service
representative.