1260 Infinity II Flexible Pump User Manual 229
8Maintenance
Prepare the Pump Module for Transport
Prepare the Pump Module for Transport
When
If the module shall be transported or shipped.
Parts required p/n Description
9301-0411 Syringe; Plastic
9301-1337 Syringe adapter
G7104-44000 Transport protection foam (G7104A/C)
G7131-98000 Transport protection foam (G7131A/C)
CAUTION
Mechanical damage
 For shipping the module, insert the Protective Foam to protect the module
from mechanical damage.
 Be careful not to damage tubing or capillary connections while inserting the
module in the Protective Foam.
1 Flush all solvent channels with isopropanol.
2 Remove solvent inlet tubes from solvent reservoirs and
tubing clips at other modules.
3 Remove tubings between the seal wash function and
solvent bottle/waste.
4 Turn off the pump.
5 Remove cable and capillary connections to other
modules.
6 Remove the waste tube.
7 Remove the module from the stack.
8 Disconnect the degasser outlet tubings at the MCGV
one after another. Use a syringe for removing liquid
from the degasser and the solvent tubings.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7104C and is the answer not in the manual?

Agilent Technologies G7104C Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7104C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals