1260 Infinity II Flexible Pump User Manual 167
8Maintenance
Install the Jet Weaver
Install the Jet Weaver
When
Optional Jet Weavers are available for applications which require highest mixing performance, see
chapter Optimizing Performance.
Tools required p/n Description
8710-0510 Open-end wrench 1/4 — 5/16 inch
8710-0899 Screwdriver Pozidrive Shaft
5023-2502 Hex driver SW-6.35, slitted
Parts required # p/n Description
1 G4204-68035 Jet Weaver Mixer Kit V35 Quat/Flex
1 G7131-68035 Jet Weaver Mixer Kit V35 Flex Bio-Compatible
1 G4204-68100 Jet Weaver Mixer Kit V100 Quat/Flex
1 G7131-68100 Jet Weaver Mixer Kit V100 Flex Bio-Compatible
1 G4204-68380 Jet Weaver Mixer Kit V380 Quat/Flex
each containing two 5500-1253
1 G7131-68380 Jet Weaver Mixer Kit V380 Flex Bio-Compatible
each containing two 5004-0041
2 5500-1253 Capillary ST 0.17 mm x 130 mm SX/S
Jet Weaver to Multipurpose Valve
2 5004-0041 Capillary MP35N 0.17 mm x 130 mm SI/SX
Preparations
Switch off the pump at the main power switch
1 Open the screw of the Jet Weaver metal lid. 2 Remove the metal lid by lifting it up (1) and pulling it out
of the front panel (2).