7 Error Information
Pump Error Messages
Reading of multipurpose valve tag failed
Error ID: 29240
Reading the multipurpose valve tag failed.
• Parameter: None
Pump drive encoder rollover
Error ID: 29232
Invalid pump drive encoder signals have been detected.
• Parameter: 1 – 2 referring to pump drive
Drive position limit
Error ID: 29234
Internal error.
• Parameter: 1 – 4 referring to pump drive
Probable cause Suggested actions
1 Reading of multipurpose valve tag failed.
Check cable connection.
2 Multipurpose valve head tag defect or empty.
Replace multipurpose valve head.
3 Multipurpose valve tag reader is defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Pump drive encoder is defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Internal error.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7104C and is the answer not in the manual?

Agilent Technologies G7104C Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7104C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals