1260 Infinity II Flexible Pump User Manual 137
7 Error Information
Pump Error Messages
Overcurrent of pump drive
Error ID: 29202
The current consumption of the pump drive is too high.
• Parameter: 1 – 2 referring to pump drive
Deliver underrun
Error ID: 29233
Internal error.
• Parameter: None
Probable cause Suggested actions
1 Blockage of system before pressure sensor.
Check for blockage of e.g. outlet valve filter frit,
Multipurpose Valve, heat exchanger.
2 Drive motor defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Internal error.
Please contact your Agilent service
representative.
2 Firmware issue
Use a minimum firmware revision of B.06.55