8Maintenance
Replace the Seal Wash Sensor
1260 Infinity II Flexible Pump User Manual 226
Replace the Seal Wash Sensor
Parts required p/n Description
5067-6172 Seal Wash Sensor Pre Assembly
Preparations
• Shut-down the pump.
• Remove the seal wash tubing from the seal wash solvent bottle.
1 Remove the solvent inlet and outlet tubes from the seal
wash sensor.
2 Press at the sides of the seal wash sensor.
3 Remove the seal wash sensor from the module chassis. 4 Remove the cable.