2 Site Requirements and Specifications
Physical Specifications
1260 Infinity II Flexible Pump User Manual 30
Physical Specifications Agilent 1260 Infinity II Bio
Flexible Pump (G7131C)
Table 2 Physical Specifications
Type Specification Comments
Weight 16.8 kg (37.0 lbs)
Dimensions
(height × width × depth)
180 x 396 x 436 mm
(7.1 x 15.6 x 17.2 inches)
Line voltage 100 – 240 V~, ± 10 % Wide-ranging capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 120 VA / 110 W
Ambient operating
temperature
4 – 55 °C (39 – 131 °F)
See footnote
1
1
At operating temperatures <15 °C the Low Temperature Environment Kit (G7131-67000) must be in-
stalled.
Ambient non-operating
temperature
-40 – 70 °C (-40 – 158 °F)
Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Non-operating altitude Up to 4600 m (15092 ft) For storing the module
Safety standards:
IEC, EN, CSA, UL
Overvoltage category II, Pollution degree 2 For indoor use only.
ISM Classification ISM Group 1 Class B According to CISPR 11