EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>G7104C

Agilent Technologies G7104C User Manual

Agilent Technologies G7104C
310 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #30 background imageLoading...
Page #30 background image
2 Site Requirements and Specifications
Physical Specifications
1260 Infinity II Flexible Pump User Manual 30
Physical Specifications Agilent 1260 Infinity II Bio
Flexible Pump (G7131C)
Table 2 Physical Specifications
Type Specification Comments
Weight 16.8 kg (37.0 lbs)
Dimensions
(height × width × depth)
180 x 396 x 436 mm
(7.1 x 15.6 x 17.2 inches)
Line voltage 100 – 240 V~, ± 10 % Wide-ranging capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 120 VA / 110 W
Ambient operating
temperature
4 – 55 °C (39 – 131 °F)
See footnote
1
1
At operating temperatures <15 °C the Low Temperature Environment Kit (G7131-67000) must be in-
stalled.
Ambient non-operating
temperature
-40 – 70 °C (-40 – 158 °F)
Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Non-operating altitude Up to 4600 m (15092 ft) For storing the module
Safety standards:
IEC, EN, CSA, UL
Overvoltage category II, Pollution degree 2 For indoor use only.
ISM Classification ISM Group 1 Class B According to CISPR 11

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7104C and is the answer not in the manual?

Agilent Technologies G7104C Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7104C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals