8Maintenance
Introduction to Maintenance
1260 Infinity II Flexible Pump User Manual 152
Figure 20 Overview of maintenance parts for the Bio Flexible Pump
Recommended Interval for Preventive Maintenance
The recommended interval for preventive maintenance is:
100 L (150 L for Long Life Technology) or 1 year (whichever comes first).
This recommendation is valid for LC instruments on which “typical” applications
are running.
A “typical” application can be characterized as follows:
pressure range 100 800 bar,
flow rates 0.5 3.5 mL/min,
typical solvents used in reversed phase LC.
Seal Wash Sensor
Seal Wash Pump
Jet Weaver (optional)
Multi-Channel
Gradient Valve (MCGV)
Inlet Weaver
Pump Head Assembly
Multipurpose Valve
Pressure Sensor
Filter Housing
High Pressure
Degasser