8Maintenance
Replace the Outlet Valve (G7104C)
1260 Infinity II Flexible Pump User Manual 212
9 Counter the outlet valve and tighten the lock screw of
the heat exchanger capillary with a torque wrench set to
approx. 3 Nm.
Next Steps:
10 Place the cap on the Outlet Valve.
11 Install the pump head assembly, reconnect all hydraulic
connections, and power up the pump.
12 Purge the system to remove air.
3 Nm
approx.