8Maintenance
Replace the High Pressure Outlet Filter or Filter Frit
1260 Infinity II Flexible Pump User Manual 220
3 Unscrew the high pressure outlet filter. 4 Replace the filter frit.
NOTE
Remove the old filter frit with the sharp side of the blade of a
knife.
5 Reassemble the high pressure outlet filter (approx.
16 Nm).
6 Reinstall the high pressure outlet filter to the filter
holder.
7 Mount the capillary connection to the pressure sensor
(1) and to the Multipurpose valve (2).
Frit
16 Nm
approx.