EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>G7117C

Agilent Technologies G7117C User Manual

Agilent Technologies G7117C
230 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #128 background imageLoading...
Page #128 background image
Agilent InfinityLab LC Series Diode Array Detectors User Manual 128
8 Test Functions and Calibration
Self-Test
Self-Test
The self-test runs a series of individual tests (described on the next pages), and
evaluates the results automatically. The following tests are run:
Slit Test (G7117B only)
Dark Current Test
Intensity Test
Wavelength Verification Test
ASTM Noise Test, a simplified version of the ASTM Drift and Noise Test
(without testing the Drift)
When
For complete detector check.
Parts required # Description
1 Max-Light Cartridge Cell (filled with water)
OR 1 Max-Light Cartridge Test Cell
Preparations
Lamp must be on for at least 10 minutes.
For noise test a longer warm-up time may be required (> 2 hours).
When using a Max-Light Cartridge Cell a flow rate of 0.5 mL/min with water is required.

Table of Contents

Other manuals for Agilent Technologies G7117C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7117C and is the answer not in the manual?

Agilent Technologies G7117C Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7117C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals