Agilent InfinityLab LC Series Diode Array Detectors User Manual 29
2 Site Requirements and Specifications
Physical Specifications
Physical Specifications
Table 1 Physical Specifications G4212-64050
Type Specification Comments
Weight 11.5 kg (25.4 lbs)
Dimensions
(height × width × depth)
140 x 396 x 436 mm
(5.5 x 15.6 x 17.2 inches)
Line voltage 100 – 240 V~, ± 10 % Wide-ranging capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 110 VA, 100 W
Ambient operating tempera-
ture
4 – 40 °C (39 – 104 °F)
Ambient non-operating tem-
perature
-40 – 70 °C (-40 – 158 °F)
Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Safety standards: IEC, EN,
CSA, UL
Overvoltage category II, Pollution degree 2 For indoor use only.
ISM Classification ISM Group 1 Class B According to CISPR 11