Agilent InfinityLab LC Series Diode Array Detectors User Manual 156
9 Maintenance
Overview of Maintenance
Overview of Maintenance
The following pages describe maintenance (simple repairs) of the detector that
can be carried out without opening the main cover.
Table 24 Overview of Maintenance
Procedure Typical Frequency Notes
Cleaning of module If required
Deuterium lamp
exchange
If noise and/or drift exceeds your application limits or
lamp does not ignite.
A wavelength calibration test and an
intensity test should be performed after
replacement.
Flow cell exchange If leaking or if intensity drops due to contaminated flow
cell.
A wavelength calibration test should be
performed after replacement.
Leak sensor drying If leak has occurred. Check for leaks.
Leak handling System
replacement
If broken or corroded. Check for leaks.