Agilent InfinityLab LC Series Diode Array Detectors User Manual 207
13 LAN Configuration
TCP/IP parameter configuration
TCP/IP parameter configuration
To operate properly in a network environment, the LAN interface must be
configured with valid TCP/IP network parameters. These parameters are:
IP address
Subnet Mask
Default Gateway
The TCP/IP parameters can be configured by the following methods:
by automatically requesting the parameters from a network-based DHCP
Server (using the so-called Dynamic Host Configuration Protocol). This mode
requires a LAN-onboard Module or a G1369C LAN Interface card, see “Setup
(DHCP)” on page 212
by manually setting the parameters using Telnet
by manually setting the parameters using the Local Controller
The LAN interface differentiates between several initialization modes. The
initialization mode (short form ‘init mode’) defines how to determine the active
TCP/IP parameters after power-on. The parameters may be derived non-volatile
memory or initialized with known default values. The initialization mode is
selected by the configuration switch, see Table 32 on page 209.

Table of Contents

Other manuals for Agilent Technologies G7117C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7117C and is the answer not in the manual?

Agilent Technologies G7117C Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7117C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals