Agilent InfinityLab LC Series Diode Array Detectors User Manual 151
8 Test Functions and Calibration
Dark Current Test
Test Failed
Probable cause Suggested actions
1 Defective slit assembly (stray light).
• Run the “Self-Test” on page 128.
• Run the “Slit Test (G7117B)” on page 140
(part of the “Self-Test” on page 128).
2 Defective detector main board.
Please contact your Agilent service representa-
tive.
3 Defective PDA/optical unit.
Please contact your Agilent service representa-
tive.