Agilent InfinityLab LC Series Diode Array Detectors User Manual 159
9 Maintenance
Replace the Deuterium Lamp
Replace the Deuterium Lamp
When
If noise or drift exceeds application limits or lamp does not ignite.
Tools required Description
Screwdriver POZI 1 PT3
Parts required # p/n Description
1 5190-0917 Long-life Deuterium lamp (8-pin) with RFID tag
Preparations
Turn the lamp off.
WARNING
Injury by touching hot lamp
If the detector has been in use, the lamp may be hot.
If so, wait for lamp to cool down.
NOTE
The lamp house cover includes a magnet.
1 Open the doors. 2 Locate the lamp cover.