Agilent InfinityLab LC Series Diode Array Detectors User Manual 77
5 Optimizing the Detector
Optimization Overview
Optimization Overview
Table 15 Optimization Overview
Parameter Impact
1 Selection of flow cell peak resolution versus sensitivity
Choose flow cell according to used column (“Choosing a Flow Cell” on
page 78).
2 Connection of flow cell chromatographic resolution
3 Setting the peak width (response time) peak resolution versus sensitivity ver-
sus disk space
Use peak width according “Choosing a Flow Cell” on page 78 as starting
point.
Set the peak-width close to the width of a narrow peak of interest in your
chromatogram.
4 Setting wavelength and bandwidth
Sample wavelength:
Never miss a peak by the use of a browser wavelength like 250 nm with
100 nm bandwidth.
Select specific wavelength with reduced optical bandwidth if you need
selectivity, e.g. 254.0 nm / 4 nm and 360.0 nm / 100 nm as reference
wavelength.
Set the sample wavelength to a peak or valley to get best linearity in
general; select a valley to get best linearity for high concentrations.
sensitivity versus selectivity
sensitivity versus linearity
Reference wavelength:
Select the reference wavelength with broad bandwidth (30...100 nm)
wavelength range where your analytes have little or no absorbance (e.g.
sample at 254 nm, reference at 320 nm).
Select the reference wavelength as near as possible to the UV range.
baseline drift due to RI effects.
5 Setting the slit width (G7117B only)
Use 4 nm slit for normal applications.
Use narrow slit (e.g 1 nm) if your analytes have narrow absorbance bands
and for high concentrations.
Use a wide slit (e.g. 8 nm) to detect very low concentrations.
Optimizing spectral acquisition (DAD only)
Set the spectral wavelength range (for colorless samples 190...400 nm is
sufficient).
Set step to 4 nm for normal use; set small step (and slit width) if high resolu-
tion of spectra with fine structure is wanted.
spectral resolution, sensitivity and lin-
earity.

Table of Contents

Other manuals for Agilent Technologies G7117C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7117C and is the answer not in the manual?

Agilent Technologies G7117C Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7117C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals