EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>G7117C

Agilent Technologies G7117C User Manual

Agilent Technologies G7117C
230 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #136 background imageLoading...
Page #136 background image
Agilent InfinityLab LC Series Diode Array Detectors User Manual 136
8 Test Functions and Calibration
Quick Noise Test
Test Failed
Quick Noise Test Evaluation
Probable cause Suggested actions
1 Insufficient lamp warm-up time.
Allow detector and UV-lamp turned on for at least
2 hours.
2 Absorbing solvent or air bubble in flow cell.
Ensure the flow cell is filled with water, and free
from air bubbles.
3 Dirty or contaminated flow cell.
Flush flow cell.
Clean the flow cell as described in “Clean the
Max-Light Cartridge Cell” on page 165.
4 Old UV-lamp.
Exchange the UV-lamp.

Table of Contents

Other manuals for Agilent Technologies G7117C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7117C and is the answer not in the manual?

Agilent Technologies G7117C Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7117C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals