Agilent InfinityLab LC Series Diode Array Detectors User Manual 78
5 Optimizing the Detector
Optimization Overview
Choosing a Flow Cell
Several flavors of the Max-Light Cartridge Flow Cell are available, see Table 16 on
page 78.
High Sensitivity
If higher sensitivity is necessary, the Max-Light Cartridge Cell ( 60 mm, V(σ)
4.0 µL) (G4212-60007) can be used. This cell enhances the detector by lowering
the limit of detection (LOD) by a factor of about 3 (depending on the application).
Normal Applications
The Max-Light Cartridge Cell ( 10 mm, V(σ) 1.0 µL) (G4212-60008) covers a wide
range of applications:
• all column diameter down to at least 2.1 mm ID or even less
• applications with peak dispersion (Peakwidth x flow) down to ~2 µL [example:
pw = 0.04 min at flow = 0.1 mL/min gives peak dispersion of 0.04 min x
0.1 mL/min = 0.004 mL = 4 µL]
Ultra-Low Dispersion
The Max-Light Cartridge ULD cell can be used with the G7117A DAD FS and
G7117B DAD. The cell is a requirement for the Ultra-Low Dispersion Kit solution
which currently exists as 1290 Infinity Ultra-Low Dispersion Kit (5067-5189). The
cell should be part of the ultra-low dispersion solution.
Table 16
Specifications for Max-Light Cartridge Flow Cells
Cartridge Cells • Max-Light Cartridge Cell ( 10 mm, V(σ) 1.0 µL) (G4212-60008)
• Max-Light Cartridge Cell ( 60 mm, V(σ) 4.0 µL) (G4212-60007)
• HDR Max-Light Cartridge Cell ( 3.7 mm, V(σ) 0.4 µL) (G4212-60032)
• ULD Max-Light Cartridge Cell ( 10 mm, V(σ) 0.6 µL) (G4212-60038)
• Max-Light Cartridge Test Cell (G4212-60011)
• Max-Light Cartridge Cell LSS ( 10 mm, V(σ) 1.0 µL) (G7117-60020)
Maximum pres-
sure
70 bar (1015 psi) Maximum Operating Pressure (MOP)60 bar (870 psi for LSS
Cell
1
150 bar (2175 psi) Maximum Incidential Pressure (MIP)
2
1
Maximum Operating Pressure (MOP): The maximum pressure at which the system can operate continu-
ously under normal conditions.
2
Maximum Incidential Pressure (MIP): The maximum pressure which the system can experience during a
short time.
pH range 1.0-12.5 (solvent dependent)