EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>G7117C

Agilent Technologies G7117C User Manual

Agilent Technologies G7117C
230 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #79 background imageLoading...
Page #79 background image
Agilent InfinityLab LC Series Diode Array Detectors User Manual 79
5 Optimizing the Detector
Optimization Overview
High Dynamic Range
The Max-Light Cartridge HDR cell can be used with the G7117A DAD FS and
G7117B DAD. The cell is required as a part of the High Dynamic Range (HDR)
solution.
Recommendations
For G4212-60007 and G4212-60008
For G4212-60007 and G4212-60008
The use of Peek-FS capillaries is not recommended. In combination with the SST
zero dead volume fitting (e.g. at the inlet) the capillary could break and the glass
particles could block/damage the flow cell.
NOTE
To protect the flow cell against overpressure (e.g. in systems with LC/MS) install
Inline Pressure Relief Valve Kit (G4212-68001), see “Inline Pressure Relief Valve
Kit (G4212-68001)” on page 81.

Table of Contents

Other manuals for Agilent Technologies G7117C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7117C and is the answer not in the manual?

Agilent Technologies G7117C Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7117C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals