Agilent InfinityLab LC Series Diode Array Detectors User Manual 140
8 Test Functions and Calibration
Slit Test
Slit Test
Slit Test (G7117B)
The slit test verifies correct operation of the micromechanical slit.
During the test, the slit is moved through all slit positions while the detector
monitors the lamp intensity change. When the slit position is changed, the
intensity drop (move to smaller slit) or intensity increase (move to larger slit)
must be within a defined range.
If the intensity changes are outside the expected range, the test fails.
1 Run the Slit Test with the Agilent Lab Advisor (for further information see
Online-Help of user interface).
Figure 43 Slit Test – Results
When
In case of problems.
Parts required # Description
1 Max-Light Cartridge Cell (filled with water)
OR 1 Max-Light Cartridge Test Cell
Preparations
Lamp must be on for at least 10 minutes.
When using a Max-Light Cartridge Cell a flow rate of 0.5 mL/min with water is required.

Table of Contents

Other manuals for Agilent Technologies G7117C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7117C and is the answer not in the manual?

Agilent Technologies G7117C Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7117C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals