Agilent InfinityLab LC Series Diode Array Detectors User Manual 201
12 Hardware Information
Setting the 6-bit Configuration Switch
Forced Cold Start
A forced cold start can be used to bring the module into a defined mode with
default parameter settings.
• Boot Main System / Revert to Default Data
The instrument will boot to main mode and changes to the module’s default
parameter. May be also required to load resident firmware into the module.
• Boot Resident System / Revert to Default Data
The instrument will boot to resident mode and changes to the module’s
default parameter. May be also required to load main firmware into the
module.
If you use the following switch settings and power the instrument up again, it will
start as described above.
CAUTI ON
Loss of data
Forced cold start erases all methods and data stored in the non-volatile
memory. Exceptions are calibration settings, diagnosis and repair log books
which will not be erased.
Save your methods and data before executing a forced cold start.
Table 31 Boot Resident / Forced Coldstart
SW1 SW2 SW3 SW4 SW5 SW6 Init Mode
1 0 0 0 0 0 Boot Main System / Keep Data
1 1 0 0 0 0 Boot Resident System / Keep Data
1 0 0 0 0 1 Boot Main System / Revert to Default Data
1 1 0 0 0 1 Boot Resident System / Revert to Default Data
Note: The setting '0' (down) is essential.