Agilent InfinityLab LC Series Diode Array Detectors User Manual 204
12 Hardware Information
Early Maintenance Feedback
Setting the EMF Limits
The setting of the EMF limits must be optimized over one or two maintenance
cycles. Initially the default EMF limits should be set. When instrument
performance indicates maintenance is necessary, take note of the values
displayed by the EMF counters. Enter these values (or values slightly less than the
displayed values) as EMF limits, and then reset the EMF counters to zero. The
next time the EMF counters exceed the new EMF limits, the EMF flag will be
displayed, providing a reminder that maintenance needs to be scheduled.
NOTE
This function is only available via Agilent Lab Advisor or Instant Pilot.