Agilent InfinityLab LC Series Diode Array Detectors User Manual 56
4 Preparing the Module
Setting up an Analysis
1 Open the purge valve of your pump (by turning it counterclockwise) and set
flow rate to 3 5 mL/min.
2 Flush all tubes with at least 30 mL of solvent.
3 Set flow to required value of your application and close the purge valve.
Table 6 Choice of Priming Solvents for Different Purposes
Activity Solvent Comments
After an installation
When switching between reverse
phase and normal phase (both times)
Isopropanol
Isopropanol
Best solvent to flush air out of the sys-
tem
Best solvent to flush air out of the sys-
tem
After an installation Ethanol or Methanol Alternative to Isopropanol (second
choice) if no Isopropanol is available
To clean the system when using buf-
fers
After a solvent change
Bidistilled water
Bidistilled water
Best solvent to re-dissolve buffer crys-
tals
Best solvent to re-dissolve buffer crys-
tals
After the installation of normal phase
seals (P/N 0905-1420)
Hexane + 5% Isopropanol Good wetting properties
NOTE
The pump should never be used for priming empty tubings (never let the pump
run dry). Use a syringe to draw enough solvent for completely filling the tubings
to the pump inlet before continuing to prime with the pump.
NOTE
Pump for approximately 10minutes before starting your application.

Table of Contents

Other manuals for Agilent Technologies G7117C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7117C and is the answer not in the manual?

Agilent Technologies G7117C Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7117C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals