Agilent InfinityLab LC Series Diode Array Detectors User Manual 86
5 Optimizing the Detector
Optimizing for Sensitivity, Selectivity, Linearity and Dispersion
Table 17 Peak Width — Response Time — Data Rate (G7117B)
Peak width at half
height [min]
1
1
Values in the User Interface may be rounded.
Response
[s]
Signal data
rate [Hz]
Scan data rate
[HZ]
≤126 pts/scan
Scan data rate
[HZ]
≤251 pts/scan
Scan data rate
[HZ]
≤501 pts/scan
Scan data rate
[HZ]
>501 pts/scan
< 0.00078125 0.0078125 240 240 80 40 20
> 0.00078125 0.015625 240 240 80 40 20
> 0.0015625 0.03125 160 160 80 40 20
> 0.003125 0.0625 80 80 80 40 20
> 0.00625 0.125 40 40 40 40 20
> 0.0125 0.25 20 20 20 20 20
> 0.025 0.5 10 10 10 10 10
> 0.05 1 5 5 5 5 5
> 0.1 2 2.5 2.5 2.5 2.5 2.5
> 0.2 4 1.25 1.25 1.25 1.25 1.25
> 0.4 8 0.625 0.625 0.625 0.625 0.625
> 0.85 16 0.3125 0.3125 0.3125 0.3125 0.3125
Table 18 Peak Width — Response Time — Data Rate (G7117A/G7117C)
Peak width at half
height [min]
1
Response [s] Scan data
rate[Hz] ≤251
pts/scan
Scan data
rate[Hz] ≤501
pts/scan
Scan data
rate[Hz] > 501
pts/scan
< 0.0015625 0.015625 120 120 40 20
> 0.0015625 0.03125 120 120 40 20
> 0.003125 0.0625 80 80 40 20
> 0.00625 0.125 40 40 40 20
> 0.0125 0.25 20 20 20 20
> 0.025 0.5 10 10 10 10
> 0.05 1 5 5 5 5
> 0.1 2 2.5 2.5 2.5 2.5
> 0.2 4 1.25 1.25 1.25 1.25