3Using the Module
Using the Optional Integrated Column Compartment
9 Select the column information to import from the list of columns and click OK.
The column information is imported into the Column Tag Information table.
10 Click the >> button at the top right of the Column Tag Information table to
show the hidden table columns. Use the horizontal scroll bar to access the
columns at the right of the table.
11 If all the information for the column is correct, click the Ok/Write Tag button to
write the information to the column ID tag.
The information in the ChemStation's Edit Columns table is also updated.
12 You can edit the information on the column ID tag using the ChemStation Edit
Columns table. When you have finished editing the information, repeat steps 7
to 9 to update the information in the tag.
It may take several seconds before the information appears in the Column Tag
Information table.
The characters semicolon (;), single quote (') and double quote (") are invalid for
the Column Tag Information table. If these characters are used in any field of the
table, an error is displayed, and OK/Write Tag button is disabled. You must delete
all invalid characters before you can write the data to the tag.
The Description field is limited to 32 characters in the Column Tag Information
table.
At this stage, you can assign a color to the column using the drop-down list in the
Color Code column; this information is displayed in the Visualization panel, but is
not written to the tag.