Installing the GC 1
Agilent Intuvo 9000 GC Installation 51
Vent hazardous gases to a fume hood.
Hazardous sample gases or uncombusted hydrogen gas can exit 
the GC from ECD exhaust, split vent and septum purge 
exhausts, and the TCD exhaust. Such gases must be safely 
vented in accordance with local safety procedures and 
standards.
If using an ECD, or if using hydrogen carrier gas that will be 
uncombusted, you must either safely vent the exhaust or 
operate the GC inside a fume hood. For example, if using 
hydrogen carrier gas, the GC would vent uncombusted hydrogen 
from a thermal conductivity detector (TCD) and from the inlet 
split vent and septum purge vent.
The ECD exhaust vents through a coiled tube. Connect tubing 
from the tube fitting at the end of this tubing to an exhaust hood 
via a hole in the back panel.
For a TCD, you must supply vent tubing and fittings to connect 
to the detector exhaust tube on the top of the detector. Route 
the tubing out the back of the GC, following the same path as for 
the ECD vent tubing.
The other detectors (FID and FPD) combust any hydrogen 
carrier gas. 
Install FID vent chimney
If FID combustion will create corrosive gases, such as the HCl 
produced during the combustion of methylene chloride solvent, 
install the FID vent chimney. Connect the exhaust to a fume 
hood as needed.
Detector exhaust vents here
Combustion of methylene chloride in the FID creates corrosive 
gases that can damage the GC. If FID combustion will create 
corrosive gases, install the FID vent chimney.