100 Agilent Intuvo 9000 GC Installation
B Cabling Diagrams and Remote Start/Stop
Using the Remote Start/Stop Cable
Remote start/stop is used to synchronize two or more 
instruments. For example, you might connect an integrator and 
the GC so that the [Start]/[Stop] buttons on either instrument 
control both of them. You can synchronize a maximum of ten 
instruments using Remote cables.
Connecting Agilent products
If connecting two Agilent products with Remote cables, the 
sending and receiving circuits will be compatible—just plug in 
both ends of the cable.
Connecting non-Agilent products
If connecting to a non-Agilent product, the following paragraphs 
contain information you will need to ensure compatibility.
APG Remote signal electrical specifications
The APG signals are a modified open collector type. The signal 
levels are generally TTL levels (low voltage is logic zero, high 
voltage is logic one) but the open circuit voltage will be between 
2.5 and 3.7 V. The typical voltage is 3 V. A voltage over 2.2 V will 
be interpreted as a high logic state while a voltage below 0.4 V 
will be interpreted as a low logic state. These levels provide 
some margin over the specifications of the devices used.
The pull-up resistance, connected to the open-circuit voltage, is 
in the range of about 1 kOhms to 1.5 kOhms. For a logic-low 
state, for a single device on the bus, the minimum current you 
must be able to sink is 3.3 mA. Since devices are connected in 
parallel, when you have multiple devices this minimum current 
must be multiplied by the number of devices attached on the 
bus. The maximum voltage for a low-input state is 0.4 V.
The bus is passively pulled high. Leakage current out of a port 
must be less than 0.2 mA to keep the voltage from being pulled 
lower than 2.2 V. Higher leakage current may cause the state to 
be interpreted as a low. 
Over-voltage protection: APG Remote connections are clamped 
by a Zener diode to 5.6 V. Exceeding this voltage will damage the 
circuit (GC logic board).