EasyManua.ls Logo

Agilent Technologies Intuvo 9000 - Condition the Column and Bake out the Detector

Agilent Technologies Intuvo 9000
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installing the GC 1
Agilent Intuvo 9000 GC Installation 85
Condition the column and bake out the detector.
Before use, the column must be conditioned to remove any
contaminants.
1 Turn on the carrier gas.
2 Perform a leak check. Touch Diagnostics and navigate to the
leak check.
3 Refer to the checkout method for the detector. See the
Operation Manual. Note the oven temperature, average
velocity or flow, and so on.
4 Set the oven temperature and inlet flow conditions specified
for conditioning the column. Touch Method, select Active
method, then touch Edit to access current GC temperatures
and flows.
Set the column temperature to 20 °C higher than the
highest checkout method temperature (do not exceed the
column maximum temperature). For example, for FID
checkout set the column temperature to 210 °C.
Set the detector temperature to 20 °C higher than the
checkout method temperature. For example, for FID
checkout set the detector temperature to 320 °C.
Turn the detector gases on.
Light the flame, if appropriate.
5 Condition for the time given in the column’s instructions.
6 Cool the oven.
7 Leave the carrier gas on. If using a flammable carrier gas
(hydrogen), continue to vent the exhausts to a fume hood.

Table of Contents

Other manuals for Agilent Technologies Intuvo 9000

Related product manuals