Testing
M3860A and M3861A H
EARTSTREAM FR2 Defibrillator 4-9
4
Screen contrast can be adjusted during the battery insertion selftest by using
the Option buttons.
If battery power is low, replace the battery. If a previous selftest has failed, the
screen displays a message that the H
EARTSTREAM FR2 must pass a selftest
before being used.
It is recommended that you always have a spare battery available. However, if
a screen display prompts you to replace the battery or the Status Indicator
shows a flashing red
X, but you do not have a spare battery, you can continue
to use the H
EARTSTREAM FR2 until the battery is completely depleted. This
may be necessary in an emergency.
During the first part of the selftest, the screen displays a bar that fills in as the
test continues. When that part of the test is finished, the H
EARTSTREAM FR2
beeps. The results of the selftest are automatically recorded on the data card
while the tests are running, if a data card was inserted in the FR2 prior to
installing the battery.
If the first part of the selftest fails:
The screen displays a message that the selftest has failed. After a short
time, an error code is displayed. Write down the error code and contact
Agilent Technologies for technical service.
The Status Indicator shows a flashing or solid red X.
Replace the battery with a new battery and repeat the test. If the second
selftest fails, do not use the H
EARTSTREAM FR2.
If the first part of the selftest passes:
The screen displays a message that the selftest passed then begins the
interactive part of the test.
The Status Indicator shows a flashing black hourglass.
The interactive part of the selftest requires you to respond to prompts in order
to make sure the display, buttons, lights, and speaker on the H
EARTSTREAM
FR2 are working properly.
Screen prompts guide you through a series of steps in the interactive part of
the selftest. Some ask you to observe that a feature of the H
EARTSTREAM FR2
works properly. Others ask you to take certain actions — for example, to press
RUN SELFTEST
REVIEW INCIDENT
NO DATA CARD
LOW BATTERY
NEXT
SELFTEST MUST
PASS BEFORE USE
SELFTEST FAILED
REV: 01EM30 0201
NOT READY FOR USE
SN 000000001
SELFTEST PASSED
REV: 01E30 0201
NO DATA CARD
SN 000000001
IN EMERGENCY
PRESS OFF TO QUIT

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies M3860A and is the answer not in the manual?

Agilent Technologies M3860A Specifications

General IconGeneral
BrandAgilent Technologies
ModelM3860A
CategoryMedical Equipment
LanguageEnglish

Related product manuals