Testing
M3860A and M3861A H
EARTSTREAM FR2 Defibrillator 4-11
4
Lights Screen prompt: CHECK SHOCK BUTTON LIGHT AND PADS CONNECTOR
LIGHT
Check that the lights come on.
Option Buttons Screen prompt: PRESS THE OPTION BUTTONS
Press the upper and lower Option buttons and listen for a beep to confirm each
press. Look at the screen to be sure the button presses have been verified.
Display Screen Screen prompt: CHECK DISPLAY. ADJUST CONTRAST IF NEEDED
Check the test pattern displayed on the screen. Adjust the contrast if desired using
the Option buttons.
NOTE: Screen contrast can be adjusted at any time during the battery insertion
selftest.
Shock Button Screen prompt: PRESS THE SHOCK BUTTON
Press the Shock button and listen for a beep to confirm the press.
No shock will be delivered when you press the Shock button during the test.
Look at the screen to be sure the button press has been verified.
On/Off Button Screen prompt: PRESS THE ON/OFF BUTTON
Press the On/Off button and listen for a beep to confirm press.
Look at the screen to be sure the button press has been verified.
The screen then displays a message that the test is complete.
FEATURE TE S T D E S C R IP T I O N

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies M3860A and is the answer not in the manual?

Agilent Technologies M3860A Specifications

General IconGeneral
BrandAgilent Technologies
ModelM3860A
CategoryMedical Equipment
LanguageEnglish

Related product manuals