EasyManua.ls Logo

Agilent Technologies N1913a - Figure 10-5 U2000 Series Settling Time with Auto-Filter; Table 10-9 U2000 Series Power Sensorssettling Time

Agilent Technologies N1913a
263 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications and Characteristics 10
Agilent N1913/1914A EPM Series Power Meters User’s Guide 231
For Agilent U2000 Series power sensors
In FAST mode (using Free Run trigger), for a 10 dB decreasing power
step, the settling time is 25 ms
2
.
Table 1 0 -9 U2000 Series power sensors’settling time
Figure 10-5 U2000 Series Settling Time with Auto-Filter
Number of Averages12481632641282565121024
Settling Time
1
(s)
(Normal Mode)
1 Manual filter, 10 dB decreasing power step (not across the switching point)
2 When a power step crosses the auto-range switch point of the sensor, add 25 ms.
0.045 0.09 0.17 0.34 0.66 1.3 2.6 5.2 10.4 20.9 41.9
Settling Time
1
(s)
(X2 Mode)
0.042 0.05 0.09 0.17 0.34 0.66 1.3 2.6 5.2 10.4 20.9
-50 dBm
+10 dBm
+2 dBm
-40 dBm
45 ms
90 ms
2.6 s
2.7 s
460 ms
2.8 s
39 s
42 s
42 s
-4 dBm
-10 dBm
-20 dBm
-30 dBm
Normal
Mode
45 ms
82 ms
1.3 s
1.5 s
450 ms
1.6 s
20 s
24 s
24 s
X2
Mode
Maximum sensor
power within a range
U2000/1/2/4A
U2000/1/2H
U2000/1B
+20 dBm
+40 dBm
+12 dBm
+32 dBm
+6 dBm
+26 dBm
0 dBm
-10 dBm
-20 dBm
-30 dBm
-40 dBm
+20 dBm
+10 dBm
0 dBm
-10 dBm
-20 dBm
Minimum sensor
power within a range
Sensor
Dynamic
Range
Ty p ic a l
Settling
Times

Table of Contents

Other manuals for Agilent Technologies N1913a

Related product manuals