10 Specifications and Characteristics
234 Agilent N1913/1914A EPM Series Power Meters User’s Guide
1 mW Power Reference
NOTE
The 1 mW Power Reference is provided for calibration of the E-Series, 8480 Series, and
N8480 Series power sensors.
Power Output: 1.00 mW (0.0 dBm)
Factory set to ±0.4% traceable to the National Physical
Laboratories (NPL), UK
Accuracy: ±1.2% (0 - 55 ºC)
±0.4% (25 ±10 ºC)
Frequency: 50 MHz nominal
SWR: 1.08 (0 - 55 ºC)
1.05 (typical)
Connector Type: Type N (f), 50 W