EasyManua.ls Logo

Agilent Technologies N1913a - Page 83

Agilent Technologies N1913a
263 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
General Power Meter Functions 2
N1913/1914A EPM Series Power Meters User’s Guide 59
Frequency Sweep Mode
Frequency sweep is generally used in a frequency response calibration
system where the amplitude is fixed, and the frequency of the power
source signal is swept. This mode can be used to determine the frequency
response of a device under test.
Procedures
1 Connect sensor to a power source.
2 Connect power meter TRIG OUT to power source TRIG IN using a
BNC cable. The same connection is applied for power source TRIG OUT
to power meter TRIG IN. See Figure 2- 19.
3 Press . The Channel Setup displays as shown in Figure 2-20.
4 Press . The Trigger menu is displayed.
5 Press softkey to configure a trigger.
6 Select either or .
7 Press to configure the remaining trigger parameters. The trigger
Settings menu consists of two pages. Figure 2- 21 shows page 1 and
Figure 2- 22 shows page 2.
8 Press and will be automatically enabled.
9 Press , and to allow a TTL level high to be produced
at the rear panel TRIG OUT BNC port when the meter is triggered upon a
completed measurement.
NOTE
This feature is only available when N8480 Series, E-Series E4410, or E-Series E9300 sensor
is connected.
NOTE
When N8480 Series, E-Series E4410, E-Series E9300 or U2000 Series sensor is connected,
Sensor Mode is set to AVG only by default.
Acqn
Sing Trig
Cont Trig
Settings
Source
Ext
1 of 2
Output
On

Table of Contents

Other manuals for Agilent Technologies N1913a

Related product manuals