EasyManuals Logo

Agilent Technologies U1253B User's And Service Guide

Agilent Technologies U1253B
221 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #187 background imageLoading...
Page #187 background image
Performance Tests and Calibration 6
U1253B User’s and Service Guide 161
AC mA/AC A 50 mA 3.000 mA (1 kHz) 0.9 to 1.1 × reference input value
30.000 mA (1 kHz) 0.9 to 1.1 × reference input value
500 mA 30.00 mA (1 kHz) 0.9 to 1.1 × reference input value
30.000 mA (1 kHz) 0.9 to 1.1 × reference input value
5 A 0.3000 A (1 kHz) 0.9 to 1.1 × reference input value
3.0000 A (1 kHz) 0.9 to 1.1 × reference input value
10 A 0.3000 A (1 kHz) 0.9 to 1.1 × reference input value
10.000 A (1 kHz) 0.9 to 1.1 × reference input value
Capacitance Open OPEN Open terminals
10 nF 3.000 nF 0.9 to 1.1 × reference input value
10.000 nF 0.9 to 1.1 × reference input value
100 nF 10.00 nF 0.9 to 1.1 × reference input value
100.00 nF 0.9 to 1.1 × reference input value
1000 nF 100.0 nF 0.9 to 1.1 × reference input value
1000.0 nF 0.9 to 1.1 × reference input value
10 μF 10.000 μF 0.9 to 1.1 × reference input value
100 μF 100.00 μF 0.9 to 1.1 × reference input value
1000 μF 1000.0 μF 0.9 to 1.1 × reference input value
10 mF 10.000 mF 0.9 to 1.1 × reference input value
Table 6- 3 Valid adjustment reference input values
Function Range Reference input value Valid range for reference input

Table of Contents

Other manuals for Agilent Technologies U1253B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1253B and is the answer not in the manual?

Agilent Technologies U1253B Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1253B
CategoryMultimeter
LanguageEnglish

Related product manuals