Dansensor® LeakPointer 3/3
+
EN User Guide 37
P/N 380064-F
06/2021
...and then press the button to update the Leak Correction Factor.
The final Test hole size, Average, and Leak Correction Factor values are added to the
Error/Event Log.
14. When Leak Correction Factor has been updated, prepare one more package and make a
control measurement using the test measurement function
to confirm that the
measured hole size is approx. 100 µm (provided the delta pressure is OK).
15. Save the product.