52 User Guide EN Dansensor® LeakPointer 3/3
+
P/N 380064-F
06/2021
5. If the first custom field requires entering of, for example, Batch no., a keyboard will
appear.
Enter appropriate information and press key.
6. If another custom field is required, this could, for example, be selection between 2
predefined options, e.g. packaging machines.
Make your selection and press .
7. When display changes to...
... close the chamber and keep it closed, until an appropriate vacuum has been
established inside the chamber.
Please note that if an extended stabilisation time has been set ( Settings -> Device -
> Extend stabilisation time) this time starts now and only when it expires and the
measurement symbol
starts to rotate....
...you can let go of the chamber, and the measurement is being performed.
The test can be aborted anytime by pressing the button.