34 User Guide EN Dansensor® MAP Mix Provectus
P/N 300862-K
11/2019
85* Warning Device temperature very low Place unit in
temperature
above 0 °C
Allow unit to
heat up
86 Error Device temperature too low Place unit in
temperature
above 0 °C
Allow unit to
heat up
87* Warning Device temperature very high Check cooling/
filters
If error persist,
contact service
88 Error Device temperature too high Check cooling/
filters
If error persist,
contact service
95 Warning A X-gas mixer is connected, must be X-gas Check settings
Slave mixer must
have same
number of
gasses and types
as Master mixer.
96 Error Mixer high pressure alarm on gas input X Input pressure
over 10.5 bar
Adjust input
pressure to less
than 10 bar
97 Warning Pressure alarm on mixer X gas input X Check gas supply
100 Warning Cannot bridge to a mixer with older software version Use mixer with
compatible SW
version
No. Type Message User action