Dansensor® MAP Mix Provectus EN User Guide 41
P/N 300862-K
11/2019
0x0011 Mix 2 setting
(3-gas mixer
only)
1 unsigned byte
Range 0 to 100%
Only used when a 3-gas mixer is connected. In case send to a
2-gas mixer, the mixer will return a ACK, but the received
setting is ignored
For a 3-gas mixer this parameter controls the setting of
"GAS
1 INPUT" i.e. %N
2
See "Mix 1 setting" encoding column, how setting of "Mix 2
setting" will change the setting of "Mix 1 setting"
0x0020 "Mixer table 1"
offset values
(write only)
Not used in MMP
0x0030 "Mixer table 2"
offset values
(write only)
Not used in MMP
0x0040 Vacuum buffer
set pressure
(Write only)
4 byte float. Saved in non volatile memory.
Actual set point for the pressure in the vacuum buffer.
Typical 3.0 or 5.0 bars.
0x0041 Vacuum buffer
hysteresis
(Write only)
4 byte float. Saved in non volatile memory.
Typical 0.3 bar. When buffer pressure reach "Vacuum buffer
set pressure" - "Vacuum buffer hysteresis" the valve is turned
on. When the pressure reaches "Vacuum buffer set pressure"
it is turned off.
0x0042 Vacuum buffer
min on time
(times 100ms)
(Write only)
Not used in MMP, setting is ignored.
0x0043 Vacuum buffer
min off time
(times 100ms)
(Write only)
Not used in MMP, setting is ignored.
Location Register write Encoding