PN T900-013, Rev A
Specications
Model Designation and Detection Limit Specications
ta7000R Standard Model
Sample Gas Model
Lower Detection Limit
H
2
PPB CO PPB
Nitrogen (N
2
) ta7000R-N2 <0.5 <0.5
Argon (Ar) ta7000R-Ar <0.5 <0.5
Helium (He) ta7000R-He <0.5 <0.5
Oxygen (O
2
) ta7000R-O
2
<0.5 <0.5
Performance Specications
Specication Description
Accuracy, EDL Mode
(Enhanced Detection
Limit Mode)
Greater of +0.25 PPB H
2
; +0.25 PPB CO; +10 of reading
Range 0–199.9 PPB
Response Time 5 minutes to 99 % response
30 minutes to 75 % response in EDL Mode.
Response time is independent of sample concentration.
Ambient Operating
Temperature
16 °C to 32 °C (60 °F to 90 °F)
Sample Compatibility Specic models available for N
2
, Ar, He, or O
2
, or H
2
analysis.
Resolution, Display 0.1 PPB
Resolution,
Communication Ports
0.01 PPB
Specications | 5-1